Reference Intervals and Cut-off Values for Thyroid Tests on the SNIBE MAGLUMI X6

  • 2.5 Minutes to read
  • Written by Axonlab
  • Hospital

Background & Objectives

The accuracy of thyroid diagnostics depends largely on population- and method-specific reference intervals (RI). The aim of this independent study (Lapić et al., 2025) was to establish RI and diagnostic cut-offs for a comprehensive thyroid panel on the chemiluminescence immunoassay system SNIBE MAGLUMI X6.


Methodology

In a population-based study, 305 healthy adults (thyroid-healthy cohort according to strict inclusion criteria) were examined. The analysis included 11 parameters:

  • Baseline hormones: TSH, TT3, TT4, FT3, FT4
  • Special parameters: Thyroglobulin (Tg), Reverse T3 (rT3), T-Uptake
  • Autoantibodies: Anti-Tg, Anti-TPO, TRAb

Statistical evaluation was performed using the 2.5th and 97.5th percentiles (two-sided) for hormone parameters and the 95th percentile (one-sided cut-offs) for autoantibodies.


Results

Stability & Precision: The MAGLUMI system demonstrated high reproducibility across the entire panel.

Sex-specific differences: While the majority of parameters (including TSH, FT3, FT4) were independent of sex, significant differences were observed for TT3, Anti-Tg and TRAb, indicating the need for sex-specific reference limits in laboratory reporting.

Autoantibody validation: The established cut-offs for Anti-TPO and TRAb allow for a clear differentiation in the clinical diagnosis of autoimmune diseases (Hashimoto’s thyroiditis, Graves’ disease).


Conclusion for Laboratory Management

The study confirms that the MAGLUMI X series represents a precise and clinically reliable platform for modern thyroid diagnostics. The availability of validated reference intervals for 11 parameters – including less commonly assessed markers such as Reverse T3 – provides laboratories with a high level of diagnostic depth and confidence in result interpretation.


Data source:

Publication: Diagnostics (MDPI), 2025.
Reference: Lapić I, et al. "Reference Intervals and Cut-Off Values for Thyroid Tests in the Croatian Adult Population on the Snibe MAGLUMI X6 Immunoassay Analyzer."
Identification: PMID: 41008732 (Open Access)

Download study as PDF

Popular posts
  • Instructions for downloading quality control data sheets
    Jeanette Fehlmann
  • CRP – a biomarker of viral and bacterial infections
  • CRP measurement with Microsemi CRP & Microsemi CRP LC-767
    Jeanette Fehlmann
Blog categories
  • Corporate
  • Hospital
  • Pharmacy
  • Life Science
  • General Practice
  • Molecular Diagnostics
  • Microbiology

Links

  • About us
  • What we do
  • We introduce ourselves
  • Social Responsibility
  • Jobs
  • Events
  • Blog
  • Downloads
  • Download RPD CoA
  • Support & Helpline
  • GTC
  • Privacy Policy
  • Legal Notice
  • Shipping- & Payment conditions
  • Certificate
  • FAQ
  • Startup Lab Program

Go to Support Center

  • 0848 00 29 66
  • Monday - Friday
  • 8 a.m. - 5 p.m.
  • Go to Support Center

Contact

  • Axon Lab AG
  • Täfernstrasse 15
  • 5405 Baden-Dättwil
  • +41 56 484 80 80
  • info@axonlab.ch

Newsletter

All prices are in CHF plus VAT and shipping costs and possibly cash on delivery charges, unless otherwise described.